Description
Product Specifications | |
---|---|
Environmental Specifications | |
Operating Temperature | 10°C to 28°C (50°F to 82°F) |
Operating Humidity | 10% to 90% RH, non-condensing |
Operating Altitude | -30.5 m to +3,050 m (-100 ft. to +10,000 ft.) |
Storage Temperature | -18°C to +50°C (0°F to 122°F) |
Storage Humidity | 10% to 90% RH, non-condensing |
Storage Altitude | -30.5 m to +4,570 m (-100 ft. to +15,000 ft.) |
VASER and VASER Pro Amplifier Specifications | |
Dimensions (H x W x D) | 14 cm x 47.75 cm x 31 cm (5.5 in x 18.8 in x 12.2 in) |
Weight | 6.8 kg (15 lbs.) |
Input Power | 115 VAC, 60 Hz @ 300VA 230 VAC, 50 Hz @ 300VA |
Ultrasonic Operating Frequency | 36 kHz nominal |
Controls | AC power I/O Amplitude Control Continuous/VASER (pulse) Mode Selection Button Timer Reset Button VASER Test Switch Wireless Footswitch Receiver ON Switch |
Indicators | Amplitude Display (10% to 100%) Continuous/VASER Mode (C, V) Ultrasound Vibration Active (Visual) Ultrasound Vibration Time (MM:SS) |
Alarms | Fault Condition Visual Indicator Fault Condition Audible Indicator VASERsmooth™ Handpiece Connection Audible Indicator (VASER Pro Amplifier Only) |
Connections | VASER Handpiece, VASER Footswitch, AC Power Cord |
Power Cord | Hospital Grade (See AC Power Cord Specifications Table) |
VASER Footswitch, if applicable | The VASER Footswitch must be “GEM SWITCH” with the following markings: 3A, 25 VAC, 50/60 Hz IP68 minimum (watertight rating) UL registered and CE marking |
Fuse | T 3.15 A 250 V (2x) (5 mm x 20 mm) |
IEC Safety Classification | Class I, Type BF, Continuous Operation, Ordinary Equipment with Respect to Ingress of Water |